Explanation of Status Error Codes

Changes from AT3600 + ATi to AT5600 in the use of 0020 error codes

Confirm requested test conditions are achieved, as well as correct measurement

One subtle, but important, change from the AT3600 to the AT5600 is the interpretation of the Status Codes.
Most commonly, these are used to indicate tests where the requested applied voltage or current levels could not be reached to within 1% under stable conditions.

The error code “0020” indicates “The test signal could not be trimmed to required value; incorrect test parameters for the load.” 
This is analogous to ALC (Auto Level Control) on most popular LCR meters.

Consider the following example of an LS test @1 kHz 5 V RMS, with limits 10 mH +/-10% (9-11 mH)

A - AT3600 / ATi

A - AT3600 / ATi

If, due to mismatched load and signal, the desired programmed  signal is not reached the AT3600 or ATi will report an 0020 error code.

However, it will STILL PASS if the measured parameter is within your set limits.

B - AT5600

B - AT5600

The AT5600 will FAIL if the status error code is flagged.
I.e. if the desired signals not reached, even if the measured parameter is STILL a “PASS”

From feedback, we found that customers wanted positive confirmation that the requested signal in the program was achieved, before measurement, as with a traditional LCR with ALC enabled.

The simplest way to remove this error code is to reduce the requested Voltage in the program until the "0020" error is no longer reported. This confirms that the requested voltage in your program was reached during the test.

C - AT EDITOR SETTINGS

C - AT EDITOR SETTINGS

Optionally, you may wish to make the PASS FAIL judgement ONLY on the measured parameter and your chosen limits, but to ignore the 0020 error code.
This can be achieved by selecting ”Disable test failures due to ALC error” in the AT test program options

D - AT5600 WITH ALC DISABLED

D - AT5600 WITH ALC DISABLED

This option was enabled, and the test rerun.

The AT5600 now ignores the 0020 error code and makes the PASS / FAIL judgement purely on the measured values, and the limits you have set.

Requirements
AT5600 Firmware v 7.206 or later
AT Editor v 3.26.6 or later
Please see AT5600 User user manual 14.5.1.” Disable test failures due to ALC error"

Please note
The AT3600 and ATi firmware do NOT support this option.
Regardless of the option setting, the 0020 status will never be part of the PASS / FAIL criteria for these units.