AT5600 Automatic Compensation Storage

New FREE feature for AT5600 users - Saves time whilst maintaining best measurements

What is Automatic Compensation Storage?

Existing AT users will already be familiar with the Short Circuit and Open Circuit compensation feature on the AT series (and any common LCR meter.)

Compensation allows you to remove the inherent measurement offset associated with any specific AT unit and of the test fixture or cabling you are using.
This is most obviously visible on low milliohm DCR readings and interwinding capacitance readings, but it can possibly have an effect on any test.

Historically this has meant that, after any program load, the user had to short all connectors and perform short circuit compensation. then remove the shorts and perform open circuit compensation, and then proceed with normal batch testing.

We are pleased to announce that we have now implemented Automatic Compensation Storage on the AT5600.
This allows the AT5600 to store its compensation factors without the need for repeated compensation.
This will be especially useful for customers who regularly test small batches of transformers using the same fixtures, as it allows the last valid compensation to be reused.

How does Automatic Compensation Storage work?

Every time a compensation is successfully applied it is also silently stored in the AT5600.
If the same program is then re-loaded again, the last valid compensation is automatically loaded and the date and time of the stored compensation is shown on the screen. (Click picture, right)

Each AT5600 will store the last valid compensation in its memory for the last 2000 programs.
Only one stored compensation is allowed per test program.
If you wish, a new compensation can be performed at ANY time, if you prefer to remain compensating on every batch.
Click to ZOOM

General Advice

1. Compensation, like any normal test run, should only be performed (and stored) when the unit has been warmed up for at least 30 minutes and self test is confirmed to pass.

2. Compensation storage has no “memory” of the specific cables or fixture used to perform compensation.
Hence if you have two similar fixtures moving between units there may be subtle differences between them.
Cables, Clips , Kelvin blades and fixture contacts can also degrade or become dirty over time, so should be maintained, as compensation removes the characteristics of these connections.

3. Stored compensation factors are a function of the specific unit and specific fixture/cabling.
Hence there is deliberately no way to transfer a stored compensation from one AT to another.

Stored compensation will NOT be loaded if ; 

1. Any change is made to the test program between the last compensation and the next program load.
This is because any changes to test signals for frequency, or even adding extra tests, or recording them, would render the stored compensation invalid.
The AT5600 silently checksums your test program to confirm the program has not changed and that any stored compensation is still valid.

2. You have deliberately discarded a saved compensation using the “REMOVE” buttons on the compensation screen. (click picture)

3. You have not yet compensated the test program after the upgrade to firmware V1.002.000
Click to ZOOM

Questions? Comments?

If you have any questions about the above or would like help upgrading your AT5600 to take advantage of these new features, please contact us for help.
Contact

Please see AT5600 user manual 8.1.6.5 for further details.
Requires AT5600 Firmware V1.002.000 or later.
Free upgrades available from our download page.

Email Voltech
USA 1-239-437-0494
EUR 44 1235 861173
ASIA 44-1235-364-001
Site Map
Terms Of Business
Terms Of Use
Privacy
  • Terms Of Use
  • Privacy
  • Terms Of Business
  • Site Map
Contact Voltech
USA 1-239-437-0494
EUR 44 1235 861173
ASIA 44-1235-364-001
or
Email Voltech
 
Submit "Red" = required field
Please, complete all required fields.
Please, select country from the list.
Please, enter a valid Email.
Thank you for your question. We will reply to your request as soon as possible.
An error occurred.
Voltech © 2021
Voltech on LinkedIn Voltech on YouTube Voltech on Twitter
We have placed cookies on your computer to help make this website better.
Please read our Privacy Statement for further information. Otherwise, we'll assume you're OK to continue.
OK, Don't Show This Message Again