AT5600 Demo
  • Demo Home
  • AT5600 setup
  • Fixture and TX
  • At Editor setup
  • At Editor Use
  • At Server setup
  • AT Server Use
  • Next Steps

AT5600 SETUP

1. Unboxing AT5600

First remove the AT5600 from its packaging

also in the box you will find:

a. USB cable for AT Editor use
b. Ethernet cable for AT Server use.
c. Safety interlock plug
d. Power cord

2. Interlock Bypass Plug

This includes HIPOT safety tests to 5 kV AC/7 kV DC.
The AT5600 should always be used with a safety system, such as our own Light Curtain, or alternatively your own system to protect the operator from risk of shock.
This protection is enforced by the Interlock Port on the rear of the unit.
If a valid safety device is not connected to this port, then the unit will not generate any dangerous voltages

For the purposes of this demonstration we have provided an override plug that bypasses this protection.
Customers are warned that this can make the unit dangerous if transformers are touched during testing, and as such the demonstration unit should only be used with the by-pass plug fitted by qualified personnel.


To fit the Interlock bypass plug
a. disconnect all power from the unit
b. remove the two M3 screws to remove the Interlock cover plate on the rear of the unit..
c. fit the interlock bypass plug as shown.

3. Power Supply

Finally connect the IEC mains plug onto the AT5600, and power on the unit

4. AT5600 Power on Screen

The AT5600 will then display to following screen
The screen shows
Serial number
Firmware version
Calibration date
Server settings

The Green icon in the bottom left indicated that the safety interlock is closed; i.e. that the unit can generate high voltages.

If this icon is red, then the means the safety interlock is open.

If this is the case, then check the fit of the interlock plug in the rear of the unit.

Back
Next
  • Terms Of Use
  • Privacy
  • Terms Of Business
  • Site Map
Voltech © 2019
Voltech on LinkedIn Voltech on YouTube Voltech on Twitter
We have placed cookies on your computer to help make this website better.
Please read our Privacy Statement for further information. Otherwise, we'll assume you're OK to continue.
OK, Don't Show This Message Again
 
Submit "Red" = required field
Please, complete all required fields.
Please, select country from the list.
Please, enter a valid Email.
Thank you for your question. We will reply to your request as soon as possible.
An error occurred.