 AT3600 High Voltage Transformer Tester
Superior by design, the AT3600, with its patented digital measurement techniques, has revolutionized transformer testing throughout out the world. Only the AT3600 combines high and low voltage testing in one station to provide this combination of speed, accuracy, flexibility and reliability.
AT3600  From $18,360.00 Help
High Voltage Transformer Tester
Estimated Lead Time: 56 days Help
Add to Cart
AT3600 Demonstation
Rep and Self -Demo Systems Available
Estimated Lead Time: 22 days
Request

Need a rental? Click here for information.

Tests: PHAS - Interwinding Phase Test   Back To Tests

Price: $549.00
Add to Cart

Where Used

This test is useful for most types of transformer, although unusual with line frequency transformers and tests for the phase angle between windings. It is most useful for determining the phase effect of audio and telecommunications transformers when placed in a transmission line.

Measurement Conditions

If you are programming the test from a computer that is connected to the auxiliary port of the tester, you may enter the test conditions in the following way. Select the test terminals, enter the test frequency, click on the measure button and the editor will select the correct voltage and display the measured phase angle. You can then use this measured angle as a basis for setting your limits.

If your computer is not connected to the tester, you can choose the test voltage from the following table.

N.B. When selecting your energised winding, check that your transformer does not have a large step-up turns ratio between this winding and any other windings as this may cause high voltages to be present during the test. Energise the winding with the most turns if this is the case.

Impedance of Energized Windings Test Voltage
1MΩ --> 100kΩ 5V
100kΩ --> 10kΩ 5V
10kΩ --> 1kΩ 5V
1kΩ --> 100Ω 3V
100Ω --> 10Ω 0.5V
10Ω --> 1Ω 100mV
1Ω --> 100mΩ 10mV
100mΩ --> 10mΩ 1mV
Details
Overview
Benefits
Cost Savings
Tests
Fixturing
Applications
What's In the Box?
Software
 
More Information