CTY - Continuity Continuity is a test designed to ensure that the transformer is correctly seated in its fixture and that all winding termination is good.
R - Resistance R is the direct current, (DC) resistance offered by an inductor due to the resistance of the winding. This checks that the correct wire has been wound.
LS - Series Inductance 1mV to 5V @ 20Hz to 3MHz Ensures the correct core material has been used and that the number of turns is correct. Range 1nH to 1MH
QL - Quality Factor 1mV to 5V @ 20Hz to 3MHz Ensures that core material and its assembly is correct. Unit of measurement, Q. Range 0.001 to 1000
ANGL - Angle of impedance 1mV to 5V @ 20Hz to 3MHz Ensures that the core material, wire resistance, number of turns and inter-winding capacitance combine to meet design specifications. Unit of measurement: Degrees. Range -360° to +360°
LL - Leakage inductance 1mV to 5V @ 20Hz to 3MHz Ensures that windings are positioned correctly on the bobbin and that any air gap included in the core design is the correct size. Range 1nH to 1kH
C - Inter-winding capacitance 1mV to 5V @ 20Hz to 3MHz Ensures that the insulation thickness between windings is correct. Range 100fF to 1mF
TR - Turns Ratio 1mV to 5V @ 20Hz to 3MHz Ensures that the number of turns on each winding and the winding polarity meet specification.
SURG - High voltage surge testing (AT3600 only) 100V to 5kV Ensures that the insulation material around the copper wire (usually lacquer) has not been damaged during manufacture introducing the risk of an inter-winding short circuit. Unit of measurement, mV Seconds. Range 1mVs to 1kVs.
IR - Insulation Resistance 100V to 7KV@ DC Resistance tests are designed to check for poor shielding and insulation between windings.
HPAC - High Pot AC 100V to 5.5KV@ 50Hz to 1KHz Hi-pot AC tests are isolation safety tests, which test the isolation between windings and between winding and core, winding to shield.
|